Installing Dionex Devices
Dionex ESA Coulochem III Electrochemical Detector: Installation
Device Type: |
Electrochemical Detector |
Device Driver: |
Dionex ESA Modules: Coulochem III Detector |
Hardware Options |
|
Restrictions: |
The following detector functions are not supported in Chromeleon:
|
What is required?
Hardware Installation
Safety Connection: |
Caution: Dionex recommends connecting the detector to the ESA Model 582/584 HPLC pump via an event cable, so that if an error is registered in the pump, the cells will be turned off to avoid damage.
To set external events on the Model 582 HPLC pump:
This operation sets the Event 2 of the pump's Relay 2 (the orange and yellow wires) to present a contact closure in the event of a pump error. Now when a pump error occurs, the Coulochem III will turn the cells off. After the pump has been fixed, the cells can be turned back on. |
Device Connection: |
First, install the Chromeleon software, and then connect the instrument to the USB port on the Chromeleon server PC.
|
Software Installation
In the Chromeleon Server Configuration, add the driver to a timebase. For details, see Adding, Configuring or Deleting Devices. After you add the driver, the configuration dialog opens automatically.
On each properties tab page, select the desired settings. For information about the settings on a page, click the Help button on a page or press F1.
Further Information
For detailed installation instructions, refer to the operating instructions for the instrument.
For troubleshooting information, refer to Troubleshooting for USB and TCP/IP Connections.
For an overview of the different ESA devices, refer to ESA Overview.